Antidiagonal

From NMR Wiki

(Difference between revisions)
Jump to: navigation, search
Current revision (20:39, 27 January 2009) (edit) (undo)
 
Line 8: Line 8:
[[Category:Artefacts in NMR data]]
[[Category:Artefacts in NMR data]]
 +
[[Category:Fixme]]

Current revision

Antidiagonal 1, when can we see it in 2D NOESY? What cause it? 2, how to get rid of it from the NOESY?

Daniel, maybe you can upload an image and insert it here?

It is possible that you've recorded only one dataset of the two required by the quad detection in idirect dimension to discriminate frequency. Also it is possible that you have chosen wrong parameters for processing - States-TPPI instead of Echo-Antiecho or something like that. Evgeny Fadeev 11:26, 13 May 2008 (MDT)