Avance II

From NMR Wiki

(Difference between revisions)
Jump to: navigation, search
(Specifications)
(Specifications)
Line 8: Line 8:
-
Resolution of the frequency generation is <0.005 (<0.006) Hz and <0.01 deg for its phase. Minimum time for phase and amplitude seting is 50 ns and for frequency seting is 100 ns. Time for the frequency and phase real switchings is 300ns for frequency and phase, <100ns for amplitude seting. Timing resolution for operations 25 ns, [[system clock]] timing is 12.5 ns. [[Dynamic range]] of ADC up to 22 [[bit]]s. Frequency stability <1X10<sup>-9</sup> per day.
+
Resolution of the frequency generation is <0.005 (<0.006) Hz and <0.01 deg for its phase. Minimum time for phase and amplitude setting is 50 ns and for frequency setting is 100 ns. Time for the frequency and phase real switchings is 300ns for frequency and phase, <100ns for amplitude seting. Timing resolution for operations 25 ns, [[system clock]] timing is 12.5 ns. [[Dynamic range]] of ADC up to 22 [[bit]]s. Frequency stability <1X10<sup>-9</sup> per day.
== Architecture ==
== Architecture ==

Revision as of 13:51, 17 October 2009

Avance II MicroBay
Avance II MicroBay

Contents

General description

Avance II is series of Bruker BioSpin Group produced Avance NMR spectrometers (consoles) with up to 8 RF (Radio Frequency) channels, plus lock channel. This console has a improved version Avance II+ also. It is/was produced for NMR spectrometers from 200 to 900 (950) MHz. Around 2006-2007 (?) y. it was replaced with Avance III series of consoles. The main difference of Avance II from earlier Avance series of spectrometer consoles is introduction of the DRU (Digital Receiving Unit) and the more extensive integration of different spectrometer functions into one physical block (device). Together with the application of DRU, instead of the separate ADC like were SADC, HADC or FADC the new RXAD board which integrates together receiver and ADC was also introduced. Avance II uses oversampling for the signal analog-digital conversion, and the digital lock system working on the same referencing frequency with which quadrature detector of observe channel is feeded. This allows to get spectrum reference frequency on the basis of lock signal and get correct chemical shifts scale for spectrum automatically. The console is controlled by the Bruker TopSpin software (not supported by XwinNMR).

Specifications

Resolution of the frequency generation is <0.005 (<0.006) Hz and <0.01 deg for its phase. Minimum time for phase and amplitude setting is 50 ns and for frequency setting is 100 ns. Time for the frequency and phase real switchings is 300ns for frequency and phase, <100ns for amplitude seting. Timing resolution for operations 25 ns, system clock timing is 12.5 ns. Dynamic range of ADC up to 22 bits. Frequency stability <1X10-9 per day.

Architecture

Avance II console is implemented in three different chassies: MicroBay (only 200-400 MHz and up to 3 channel configuration), OneBay (up to 4 channels and 600 MHz?) and TwoBay (Up to 8 cannels and 900 (950) MHz, and for MAS). The main parts of the Avance II console are AQS (AQuisition System) or newer version AQS/2 and BSMS/2 (Bruker Smart Magnet System). In TwoBay configuration two AQS-s modules can be installed simultaneously which is used for additional RF channels.

In AQS or AQS/2 are located the (main) devices (blocks): DRU or DRU-E, RXAD (Receiver with integrated Analog to Digital converter), REF or REF/2 (REFerence Unit), SGU (Signal Generation Unit), FCU3 or FCU4 (Frequency Control Unit), TCU3 (Timing Control Unit), GCU (Gradient Control Unit), CCU10 or CCU11 (Communication Control Unit), PSM1 or 2 (Power Supply Module). AQS/2 can also contain internal amplifiers (BLA2BB, BLAX300, only for MicroBay configuration), preamplifiers (PREAMP XBB and PREAMP 1H2H, only for MicroBay configuration), and ROUTER3/5 (Router). AQS/2 can contain more than one of each of those units.

BSMS/2 contains: ELCB (Extended Lock Control Board) or CPU/3 (Central Processing Unit) plus LCB (Lock Control Board), SLCB (Sample Level Control Board), GAB (Gradient Amplifier Board), PSB (Power Supply Board), SCB13 or SCB20 (Shim Control Boards), BVT (Bruker Variable Temperature), L-TX (Lock Transmitter), L-RX (Lock Receiver), 2H-TX (Lockswitch), PNK (console pneumatics).

Additionally, depending on configuration, Avance II console chassis can be configured with various power amplifiers, differnt VTU (Variable Temberature Unit), MAS devices, etc.

To the console external devices, like HPPR (High Power Preamplifier) or HPPR/2, KBC (keyboard), PC (computer) and others, are and can be connected.

Maintenance

Bugs

1) There are observed problems with locking and lock signal displaing, when the system is up a long time. The BSMS or console reboot solves this problem.

2) In the case of longtime experiments with rapid pulsing, there are observed problems with the communication with AQS. During of the aquisition the realtime FID and spectrum can't be monitored on screen after some time (after around 1-2 million scans) or the communication fails completely and aguisition can´t be started at all. The AQS or Console reboot solves this problem.