Setting up shaped pulses

From NMR Wiki

Revision as of 20:07, 29 March 2010 by Evgeny Fadeev (Talk | contribs)
Jump to: navigation, search

Bruker

On Bruker systems the statement for a shaped pulse is similar to:

(p3:sp3 ph3):f3

Here it is intended to apply to channel f3 (channel 3) a pulse of duration p3 microseconds using phase table ph3.

Usage of :sp3 in the statement implies these things:

  • the pulse is shaped
  • sp3 is the variable with a value of maximum power for the pulse
  • variable spnam3 contains the (string) name of the waveform file
  • any power value for channel f3 set before is ignored (???) and sp3 is used instead

Varian

explain how to set up shaped pulses on Varian systems

Links

related question on the forum