T1 measurement by saturation recovery

From NMR Wiki

Revision as of 06:04, 6 February 2009 by Evgeny Fadeev (Talk | contribs)
(diff) ←Older revision | Current revision (diff) | Newer revision→ (diff)
Jump to: navigation, search

Saturation Recovery

In the Saturation Recovery experiment, the time it takes for the spins to recover from a saturated state is measured. If one has a probehead where the use of gradients is possible, this is quite easy:

saturation recovery pulse sequence for T1 measurement using pulsed field gradient

However without the possibility to use gradients, a set of several pulses in a short timespan (often with decreasing delays between the pulses in this pulsetrain) can be used to saturate the spin system:

saturation recovery pulse sequence for T1 measurement using saturating pulse train