NANUC air drying setup
From NMR Wiki
This setup is used for drying VT air at NANUC
Balston PSA dryer (model 75-20) --> two ballast tanks (240L each). Ballast tanks are used to even out concentration of O2 during the operation cycle.
Dryer has the following specifications:
- 8.3 SCFM maximum flow rate at –73 Deg C dewpoint at 80psig
- Inlet air temperature: 40F to 78F
- Inlet pressure min/max: 60 psig to 125 psig