Ice in my dewar! (Liquid helium transport dewar)

From NMR Wiki

(Difference between revisions)
Jump to: navigation, search
Line 10: Line 10:
[http://andysoper.synthasite.com/resources/NMR_Links_and_Widgets/Liquid%20Helium%20Transport%20Dewar%20Up%20Close%20and%20Personal%20Version%201.2.1.doc download] (MS Word)
[http://andysoper.synthasite.com/resources/NMR_Links_and_Widgets/Liquid%20Helium%20Transport%20Dewar%20Up%20Close%20and%20Personal%20Version%201.2.1.doc download] (MS Word)
-
[[Category:Categorize]]
+
[[Category:Cryogenics]]

Revision as of 15:56, 12 March 2009

Here is a solution to an iced-up Liquid helium Transport Dewar. This is of special interest to sites which are far from their liquid helium supplier, and wh perform their own liquid helium fils.

Scenario:

The dewar will not pressurize, or pressurizes reluctantly. The dewar will not vent via the regular vent port. Some of the safety valves do not vent any gas.