Current revision
Instrument
INOVA 400 with VnmrJ 2.1B
Problem Description
"Number of points ## less then np" error message where ## is two less than np when starting acquisition with dsp='y'. Normal acquisition with dsp='n'.
Solution
Removed ADC board from console, removed and re-seated DSP mezzanine board on ADC board, and put ADC back into console.