X-nucleus pulse calibration on decoupler2 (G. Montelione, Rutgers Univ.)
From NMR Wiki
(Difference between revisions)
Line 5: | Line 5: | ||
|console=Unity | |console=Unity | ||
|field=500 | |field=500 | ||
+ | |dim=1 | ||
+ | |nuclei=HX | ||
+ | |observe=H | ||
}} | }} | ||
[[Category:Calibration]] | [[Category:Calibration]] | ||
- | + | [[Category:Heteronuclear]] | |
- | [[Category:Heteronuclear]] | + | [[Category:Varian]] |
Revision as of 21:09, 5 October 2008
This page is a part of the user-editable NMR Wiki pulse sequence database.
Disclaimer: You are solely responsible for correct use of these materials on your instrument.
search database | see edit instructions | see deposition instructions | report suggestions |
Name: hetcal2 | ||
Synopsis: calibration of 90o pulse on X-nucleus | ||
| ||
Copyright: G.T. Montelione, Protein NMR Laboratory - Center for Advanced Biotechnology and Medicine, Rutgers University | ||
Console: Varian, Inc. Unity, 500 MHz | ||
Console requirements: four RF channels (applies to some CABM sequences) | ||
Source: CABM Pulse Sequence Library | ||
Download (external links):readme pulse sequence parameters (tar) |